Fisheries and Aquatic Sciences
BioMed Central
Research Article

Method development and validation for dieckol in the standardization of phlorotannin preparations

Jiyoung Kim1, Minyoung Um1, Hyejin Yang1, Inho Kim1, Changho Lee1, Yuntai Kim1, Minseok Yoon1, Youngkyoung Kim2, Jimi Kim2, Suengmok Cho1,*
1Korea Food Research Institute463-746SungnamKorea
2Korea Health Supplements Institute463-400SungnamKorea
*82-31-709-9246smcho@kfri.re.kr

© Kim et al. 2016. Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Received: Feb 2, 2016; Accepted: Feb 16, 2016

Published Online: Mar 11, 2016

Abstract

Phlorotannins are reported to have diverse biological properties. However, no analytical methods for the standardization of phlorotannin preparations have been reported. Herein, we developed and validated an analytical method for the determination of dieckol in phlorotannin extracts (PRT) using high-performance liquid chromatography (HPLC). The optimum HPLC conditions consisted of a Supelco Discovery C18 column stationary phase, a mobile phase (A: 15 % HPLC grade methanol in deionized water, B: methanol), UV detection at 230 nm, and a flow rate of 0.7 mL/min. The optimized chromatographic conditions were validated and exhibited good specificity and linearity (R2 > 0.9994–1.0000). The recoveries were in the range of 100.9–102.3 %. The method had good intermediate (%RSD 1.2) and intra-day (%RSD 0.4–1.7) assay precisions. This HPLC method had good accuracy and quality in the determination of dieckol in PRT.

Keywords: Dieckol; HPLC measurement; Validation; Phlorotannins; Standardization

Background

Marine polyphenol phlorotannins are produced from secondary metabolites via the acetate-malonate pathway in brown seaweeds (Shibata et al., 2004; Isaza Martínez & Torres Castañeda 2013). They have fundamentally different structures than the polyphenols of terrestrial plants (Isaza Martínez & Torres Castañeda 2013; Shibata et al., 2002). Terrestrial plant polyphenols are based on gallic acids or flavones, whereas phlorotannins are only derived from oligomers and polymers of phloroglucinol (1,3,5-trihydroxybenzene) (Koivikko et al., 2007). Thus far, many phlorotannins, such as dieckol, eckol, triphlorethol A, bieckol, fucol, fucophlorethol, have been identified (Isaza Martínez and Torres Castañeda 2013; Kim et al., 2014; Cho et al., 2012).

Over the past 10 years, studies on the biological activities of phlorotannins have increased exponentially (Isaza Martínez & Torres Castañeda 2013). They have a wide range of biological properties, such as antioxidant (Zou et al., 2008), anti-inflammatory (Kim et al., 2009), anti-allergic (Li et al., 2008), and neuroprotective effects (Ahn et al., 2012). Recently, Cho et al. (2014) reported that a phlorotannin preparation and its constituent, eckstolonol, promoted non-rapid eye movement sleep via the benzodiazepine site of gamma-aminobutyric acid type A receptors. Therefore, phlorotannins are considered a promising material for the development of functional foods and supplements.

The standardization of phlorotannin preparations is required for the development of functional foods (Hwang et al., 2009). However, a validation for the standardization of phlorotannin products has not been reported. Dieckol is generally the most abundant compound in phlorotannin preparations, and is used as an indicator compound (Shibata et al., 2004; Cho et al., 2012; Goo et al., 2010). Therefore, we developed and validated an HPLC method for the determination of dieckol for the commercialization of phlorotannin preparations.

Methods

Reagents and materials

Phlorotannin preparations (PRT) were obtained from S&D Co., Ltd. (Cheongwon-gun, Korea). The PRT were prepared from the brown seaweed, Ecklonia cava, using a macroporous adsorption resin, as described in our previous work (Kim et al., 2014). The total phlorotannin content (TPC) of the PRT was assessed by the Folin-Ciocalteu method (Slinkard and Singleton, 1977), which determined that there were 905 mg phloroglucinol equivalents/g. Dieckol, a standard compound, was isolated using silica gel and Sephadex LH-20 column chromatography. All of the reagents used were of HPLC grade and purchased from Sigma-Aldrich Co. (St. Louis, MO, USA).

Preparation of dieckol and PRT samples

A dieckol stock solution was prepared by dissolving 5 mg in 2 mL of dimethyl sulfoxide (DMSO) and 3 mL of methanol. Analytical working solutions were prepared by diluting the stock solution with methanol to seven concentrations, i.e., 12.5, 25, 50, 100, 200, 300, and 400 μg/mL. The phlorotannin extract (50 mg) was added to a volumetric flask, and was then dissolved in 10 mL of DMSO and 40 mL of methanol using ultrasonication at ambient temperature. The PRT sample solution was filtered through a 0.45 μm PTFE syringe filter paper (Whatman, Maidstone, UK), and was used as a sample solution for the HPLC analysis.

Method development for dieckol

We evaluated chromatographic variables, such as flow rate, mobile phase, solvent ratio, and column oven temperature to optimize the chromatographic conditions. The determination of dieckol, an indicator compound in PRT, was performed using an Agilent system (Agilent, Palo Alto, CA, USA) consisting of an autosampler, pump, degasser, and column oven. Shideido capcellpak C18 UG (4.6 mm × 250 mm, 5 μm particle size; Shiseido, Tokyo, Japan), Cadenza C18 (4.6 mm × 150 mm, 3 μm particle size; Imtakt, Kyoto, Japan), and Supelco Discovery C18 (4.6 mm × 250 mm I.D., 5 μm particle size; Supelco, Bellefonte, PA, USA) columns were used for the separation of dieckol. The analytical conditions (mobile phase composition, flow rate, oven temperature, and injection volume) are summarized in Table 1. The temperature of the column oven was fixed at 25 °C, and the UV detection wavelength for the phlorotannins was set to 230 nm.

Table 1. Chromatographic conditions for dieckol in phlorotannin preparations

Parameters

HPLC measurement conditions

Set 1

Set 2

Set 3

Solutions

A: 15 vol% MeOH in deionized water, B: methanol

Column

Shideido capcellpak C18

Cadenza C18

Supelco Discovery C18

Mobile phase (A:B)

90:10

90:10

85:15

Flow rate (mL/min)

0.8

0.8

0.7

Oven temperature (°C)

35

35

25

Injection volume (μL)

10

10

5

Elution time (min)

60

60

55

Download Excel Table

Method validation

The method was validated in terms of its specificity, linearity, accuracy, recovery, and precision according to the guidelines of the International Conference on the Harmonization of Technical Requirements for Registration of Pharmaceuticals for Human Use (ICH) (Jeong et al., 2013; International Conference on Harmonization ICH 1997).

Linearity

Linearity was tested at seven concentrations, i.e., 12.5, 25, 50, 100, 200, 300, 400 μg/mL. The linear regression equation was calculated from the calibration curve as Y = Ax + B, where A and B are the slope and intercept of the calibration curve, respectively, x is the concentration of dieckol, and Y is the peak area (Goo et al., 2010; Kim et al., 2013). The correlation coefficient (R2) values were determined for the calibration curve.

Accuracy and recovery

The accuracy was measured by the recovery, which was determined by the addition of a known amount of dieckol (Ariffin et al., 2014; Moussata et al, 2014). Phlorotannin samples were mixed with three different concentrations of dieckol, and the percent recoveries were calculated using the following equation:

Recovery%=amountofdieckolrecoveredμg/mLamountofdieckoladdedμg/mL×100

Precision

The precision was determined using the intermediate (inter-day) and intra-day assays and reported as the percent relative standard deviation (%RSD) (Mittal and Parmar, 2010; Sarkar et al., 2006). The intermediate assay was performed by two different analysts over four days, and the intra-day assay was analyzed through six replicate injections of each sample solution. The %RSD was calculated using the following equation (Jeong et al., 2013):

%RSD=standarddeviationmeanmeasuredamount×100

Results and Discussion

Development of HPLC conditions for dieckol

Different chromatographic conditions (flow rate, solvent ratio, elution time, and oven temperature) were varied to optimize the HPLC analysis of dieckol in PRT (Table 1). Using the Set 1 conditions, the dieckol peak was separated from the other minor peaks (19 min), but had an unstable baseline (Fig. 1a). The Set 2 conditions did not show superior chromatographic results (data not shown). However, the Set 3 conditions had the most stable baseline near the dieckol peak (25 min), and the peaks were well resolved (Fig. 1b). The peak stabilities of dieckol and the other phlorotannin constituents in the PRT were considerably affected by the chromatographic conditions. To confirm the detection wavelength for dieckol, its UV spectrum was recorded between 200–550 nm, and it strongly absorbed at 230 nm (Fig. 1c). The PRT sample also showed same spectrum (data not shown). Henry and Alstyne (2004) reported that phlorotannins absorbed in the UV-C region between 195–265 nm. These results showed that the optimum conditions for the detection of dieckol in PRT could be found through the use of various conditions in the HPLC system.

fas-19-0-3-g1
Fig. 1. HPLC chromatograms of phlorotannin preparations detected at 230 nm using a Set 1 conditions and b Set 3 conditions. c The dieckol UV spectrum between 200–550 nm
Download Original Figure

Validation of the proposed analytical method

The validation provides reliable documentation for standardization through regulated experiments (Shabir, 2003; Epshtein 2004). According to the United States Food and Drug Administration (FDA) and Pharmacopeia, validation is a critical factor in the development of the functional food industry (Epshtein 2004; Kazakevich and Lobrutto 2007). Therefore, the validation of the analytical HPLC conditions was conducted in terms of specificity, linearity, accuracy, recovery, and precision (Satinder and Henrik 2011; Meyer, 2010).

Specificity is the ability to separate the target compound from interference in a mixture of components (Ariffin et al., 2014; Shabir, 2003). The chromatograms show the absence of any interfering peak near the retention time of dieckol (25 min) (Fig. 2a and b). This indicated that the developed chromatographic parameters (Set 3) were adequate for the acceptable resolution of dieckol in the PRT during detection (Shabir, 2003).

fas-19-0-3-g2
Fig. 2. HPLC chromatograms of a dieckol and b phlorotannin preparations for specificity verification using Set 3 conditions
Download Original Figure

Linearity is the ability to elicit results through the analysis of diverse concentrations within a given range (Shabir, 2003; Shah et al., 2014). A linear regression analysis (n = 3) was measured over seven different concentrations (12.5–400 μg/mL), and its regression equations are given in Table 2. The seven point calibration curves for dieckol were generated using the appropriate peak area ratios (data not shown). The slope ranged from 32.9–34.8 with a correlation coefficient (R2) value of 0.999. The R2 indicated an acceptable linearity over each concentration (12.5–400 μg/mL) (Shabir, 2003; Epshtein 2004). Shabir (2003) reported that R2 > 0.999 is generally considered an acceptable value for a regression line. Because each R2 was >0.998, the regression suggested good linearity between the peak areas and the compound concentrations over the tested range (Goo et al., 2010).

Table 2. Linear regression equations for dieckol in phlorotannin preparationsa

Parameters

Regression equationsb

Concentration (μg/mL)

12.5–400

Slope

34.575

32.992

34.864

Intercept

30.112

1.297

−24.646

Correlation coefficient (R2)

0.9994

1.000

0.9999

aValues were calculated using linearity analyses (n = 3)

bRegression equation is Y = Ax + B, where Y is the peak area and x is the dieckol concentration (μg/mL)

Download Excel Table

The accuracy was determined nine times at a minimum of three concentrations (13.45, 20.17, and 26.90 μg/mL) and the results are given in Table 3. The recovery of dieckol ranged from 98.8–103.1 % and the %RSD was 1.2–2.0. Shabir reported that the recovery (%) for an accuracy criterion should be 100 ± 2 % at each added concentration (Shabir, 2003). Our results showed recoveries >98 %, which indicated that the method was acceptable for the quantification of dieckol in PRT.

Table 3. Accuracy validation for dieckol in phlorotannin preparations

Phlorotannin Sample (mg)

Add dieckol (μg/mL)

Accuracya

Detected area

Total amount found (μg/mL)

Recovery (%)

Average recovery (%)a

RSD (%)

25.4

13.45

4281.71777

124.33290

101.999

101.186

2.01

26.6

4492.79346

130.42899

102.685

24.1

3959.75781

115.03436

98.873

27.1

20.17

4755.54248

138.07146

101.574

100.946

1.30

24.0

4300.78955

124.88372

101.827

25.6

4434.56396

128.74726

99.438

23.8

26.90

4506.92139

130.83702

101.807

102.372

1.29

23.2

4401.51318

127.79272

101.461

22.5

4582.42969

133.01778

103.849

RSD relative standard deviation of recovery

aValues were calculated using accuracy (n = 9) analyses

Download Excel Table

The intermediate and intra-day (repeatability) assay precision must be evaluated in terms of dieckol content (mg/g) and %RSD. Therefore, we tested the intermediate and intra-day assays through intra-lab variations, such as different days, analysts, equipment, and six replicate measurements for each sample (Shabir, 2003). As shown in Table 4, the intermediate assay results of the dieckol content and %RSD were 73.0–75.3 mg/g and 1.2, respectively. The intra-day assay results of the dieckol content and %RSD were 72.7–75.0 mg/g and 1.0, respectively. According to the International Conference on Harmonization (ICH), the precision criterion for the %RSD should be ≤2 % (International Conference on Harmonization ICH 1995; César et al., 2006). Based on our results, we confirmed that both %RSD values were <1.5 %, which indicated a good precision for the analysis of dieckol.

Table 4. Precision (Intermediate assay and Intra-day) validation for dieckol in phlorotannin preparations

Concentration (μg/mL)

Instruments

Dieckol content (mg/g)

Average content (mg/g)

SD (mg/g)

RSD (%)

Itermediate assay

94.56231

Agilent 1260

73.930

74.453

0.920

1.236

97.25632

Agilent 1260

75.057

93.42563

Agilent 1260

73.044

98.62358

Agilent 1200

75.384

Phlorotannin sample (mg)

Concentration (μg/mL)

Dieckol content (mg/g)

Average content (mg/g)

SD (mg/g)

RSD (%)

Intra-day: Agilent 1260 Infinitya

56.5

96.10084

74.1592

73.9297

0.761

1.030

59.4

101.08464

74.1968

58.3

97.25135

72.7300

56.6

95.55534

73.6080

57.9

99.64617

75.0358

52.7

89.26187

73.8485

SD standard deviation, RSD relative standard deviation

aValues were calculated using intra-day analyses (n = 6)

Download Excel Table

Conclusions

We successfully developed an HPLC method for the determination of dieckol in PRT. In addition, we systematically investigated the validation of this analysis. Because information regarding the standardization of phlorotannin preparations or brown seaweed extracts is limited, our study could be useful for the development and commercialization of phlorotannins.

Acknowledgements

This study was supported by grants from the Korea Food Research Institute (E0191402-07) and the Small and Medium Business Administration (G01981).

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

JK investigated overall HPLC analysis of dieckol from phlorotannin extracts and designed to draft the manuscript. MU participated in the design of the study and performed the statistical analysis. IK, CL and YK helped conceived to the study. HY and MY carried out isolation and purification of dieckol from phlorotannin extracts. YK and JK analyzed quantification and validation of dieckol using the HPLC instrument. SC led of this study, and helped to prepare the manuscript. All authors read and approved the final manuscript.

References

1.

Ahn BR, Moon HE, Kim HR, Jung HA, Choi JS. Neuroprotective Effect of Edible Brown Algae Eisenia bicyclis on Amyloid Beta Peptide-induce Toxicity in PC12 Cells. Arch Pharm Res. 2012; 11:1989-1998

2.

Ariffin AA, Ghazali HM, Kavousi P. Validation of a HPLC method for determination of hydroxymethylfufural in crude palm oil. Food Chem. 2014; 154:102-107

3.

César IDC, Braga FC, Soares CDV, Nunan EDA, Pianetti GA, Condessa FA, Barbosa TAF, Campos LMM. Development and validation of a RP-HPLC method for quantification of isoflavone aglycones in hydrolyzed soy dry extracts. J Chromatogr B. 2006; 836:74-78

4.

Cho S, Yang H, Jeon YJ, Lee CJ, Jin YH, Baek NI, Kim D, Kang SM, Yoon M, Yong H, Shimizu M, Han D. Phlorotannins of the edible brown seaweed Ecklonia cava Kjellman induce sleep via positive allosteric modulation of gamma-aminobutyric acid type A-benzodiazepine receptor: A novel neurological activity of seaweed polyphenols. Food Chem. 2012; 132:1133-11424

5.

Cho S, Yoon M, Pae AN, Jin YH, Cho NC, Takata Y, Urade Y, Kim S, Kim JS, Yang H, Kim J, Kim J, Han JK, Shimizu M, Huang Z. Marine polyphenol phlorotannins promote non-rapid eye movement sleep in mice via the benzodiazepine site of the GABAA receptor. Psychopharmacology. 2014; 231:2825-2837

6.

Epshtein NA. Structure of Chemical Compounds, Methods of Analysis and Process Control: Validation of HPLC Techniques for Pharmaceutical Analysis. Pharm Chem J. 2004; 38:2120-228.

7.

Goo HR, Choi JS, Na DH. Quantitative Determination of Major Phlorotannins in Ecklonia stolonifera. Arch Pharm Res. 2010; 33:539-544

8.

Henry BE, Alstyne KL. Effects of UV Radiation on Growth and Phlorotannins in Fucus gardneri (Phaeophyceae) Juveniles and Embryos. J Phycol. 2004; 40:527-533

9.

Hwang JS, Kim SJ, Kim HB. Development and Industry of Health Functional Food in Korea. Food Sci Technol Res. 2009; 15:1-4

10.

International Conference on Harmonization (ICH). Q2A: Text on Validation of Analytical Procedures. US FDA Federal Register. 1995; 60:11260.

11.

International Conference on Harmonization (ICH). Q2b: Validation of Analytical Procedures: Methodology. US FDA Federal Register. 1997; 62:27463.

12.

Isaza Martínez JH, Torres Castañeda HG. Preparation and chromatographic analysis of phlorotannins. J Chromatogr Sci. 2013; 51:825-838

13.

Jeong HC, Shim YS, Rhee YK, Choi SY, Hong HD, Chung J, Han MJ, Cho CW. Quantification of Marker Compounds in Cirsium setidens Nakai by HPLC-DAD. Food Sci Biotechnol. 2013; 22:1481-1486

14.

Kazakevich Y, Lobrutto R. HPLC for Pharmaceutical Scientists. 2007 Hoboken, New Jersey, USA: John Wiley & Sons, Ltd.

15.

Kim AR, Shin TS, Lee MS, Park JY, Park KE, Yoon NY, Kim JS, Choi JS, Jang BC, Byun DS, Park NK, Kim HR. Isolation and Identification of Phlorotannins from Ecklonia stolonifera with Antioxidant and Anti-inflammatory Properties. J Agr Food Chem. 2009; 57:3483-3489

16.

Kim SM, Kang SW, Jeon JS, Jung YJ, Kim WR, Kim CY, Um BH. Determination of major phlorotannins in Eisenia bicyclis using hydrophilic interaction chromatography: Seasonal variation and extraction characteristics. Food Chem. 2013; 138:2399-2406

17.

Kim J, Yoon M, Yang H, Jo J, Han D, Jeon Y, Cho S. Enrichment and purification of marine polyphenol phlorotannins using macroporous adsorption resins. Food Chem. 2014; 162:135-142

18.

Koivikko R, Loponen J, Pihlaja K, Jormalainen V. High-performance Liquid Chromatographic Analysis of Phlorotannins from the Brown Alga Fucus Vesiculosus. Phytochem Analysis. 2007; 18:326-332

19.

Li Y, Lee SH, Le QT, Kim MM, Kim SK. Anti-allergic Effect of Phlorotannins on Histamine Release via Binding Inhibition between lgE and FcεRI. J Agr Food Chem. 2008; 56:12073-12080

20.

Meyer VR. Practical High-Performance Liquid Chromatogaphy. 20105 Chichester, UK: John Wiley & Sons, Ltd.

21.

Mittal A, Parmar S. Development and Validation of Rapid HPLC Method for Determination of Doxofylline in Bulk Drug and Pharmaceutical Dosage Forms. J Anal Chem. 2010; 65:293-297

22.

Moussata J, Wang Z, Wang J. Development and validation of an HPLC method for the simultaneous quantification of indole-3-carbinol acetate, indole-3-carbinol, and 3,3’-diindolylmethane in mouse plasma, liver, and kidney tissues. J Chromatogr B. 2014; 958:1-9

23.

Sarkar M, Khandavilli S, Panchagnula R. Development and validation of RP-HPLC and ultraviolet spectrophotometric methods of analysis for the quantitative estimation of antiretroviral drugs in pharmaceutical dosage forms. J Chromatogr B. 2006; 830:349-354

24.

Satinder A, Henrik R. HPLC Method Development for Pharmaceuticals. 2011 London, UK: Academic. .

25.

Shabir GA. Validation of high-performance liquid chromatography method for pharmaceutical analysis understanding the differences similarities between validation requirements of the US Food and Drug Administration, the US Pharmacopeia and the International Conference on Harmonization. J Chromatogr A. 2003; 987:57-66

26.

Shah SR, Dey S, Pradhan P, Jain HK, Upadhyay UM. Method development and validation for simulataneous estimation of albendazole and praziquantel in bulk and in a synthetic mixture. J Taibah Univ Sci. 2014; 8:54-63

27.

Shibata T, Fujimoto K, Nagayama K, Yamaguchi K, Nakayama T. Inhibitory activity of brown algal phlorotannins against hyaluronidase. Int J Food Sci Tech. 2002; 37:703-709

28.

Shibata T, Kawaguchi S, Hama Y, Inagaki M, Yamaguchi K, Nakamura T. Local and chemical distribution of phlorotannins in brown algae. J Appl Phycol. 2004; 16:291-296

29.

Slinkard K, Singleton VL. Total phenol analysis: Automation and comparison with manual methods. Am J Enol Viticult. 1977; 28:49-55.

30.

Zou Y, Qian ZJ, Li Y, Kim MM, Lee SH, Kim SK. Antioxidant effects of phlorotannins isolated from Ishige okamurae in free radical mediated oxidative systems. J Agr Food Chem. 2008; 56:7001-7009